Vedere la pagina 215
B-12
TR2510 Current Probe
TR2510 Current Probe
Guidelines
To achieve the rated accuracies, follow these guidelines:
The conductor must be at a right angle to the probe.
The conductor must be centered in the probe core.
The jaw contact surfaces must be clean and properly aligned.
Specifications
Range: 0.1 to 10 Arms
Accuracy (at 60 Hz): 0.1 to 10A, ±1% reading ±0.01A
Working voltage: 600 V max.
Frequency range: 40 to 3000 Hz
Output signal: 150 mV/A
Maximum conductor size: 0.78" (20 mm)
Phase shift (at 60 Hz):
1 to 5A less than 1º
5 to 10A less than 1.5º
Operating temperature: 14 to 131 ºF (-10 to 55 ºC)
Storage temperature: -40 to 158 ºF (-40 to 70 ºC)
Dimensions: 5.47 x 2.00 x 1.18” (139 x 51 x 30 mm)
Weight: 6.5 oz. (180 g)
NOTE
Current probe TR2510 can be used interchangeably with TR2510A.